Operation
6-6
Service Manual
Figure 6-1.
Sample POST Splash Screen
If the monitoring system detects an internal problem during the POST process,
an error tone sounds and the monitoring system displays an error message.
Reference
Note:
Physiological conditions such as excessive patient movement, medical procedures, or
external agents such as dysfunctional hemoglobin, arterial dyes, low perfusion, dark
pigment, and externally applied coloring agents such as nail polish, dye, or pigmented
cream may interfere with the monitoring system’s ability to detect and display
measurements.
6.2.4
System Resets
If the monitoring system issues a system reset, based on triggering the watch-
dog timer, all temporary settings are retained. Neither factory nor institutional
default settings are impacted.
6.2.5
Automatic Shutdown and Power Off
Automatic Shutdown
When any of the following conditions are present, the monitoring system
automatically shuts down.
•
The monitoring system is running on battery power and the battery capacity
remaining reaches 0%.
•
The monitoring system has detected an internal temperature above 67 ºC or 153 ºF.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......