Power Components Replacement
Service Manual
12-23
3.
Remove the green (GND) spade connector from the J21 mating blade on the Main
PCB.
4.
Remove the green (GND) spade connector from the E1 mating blade on the power
supply PCB.
5.
Remove the two-pin blue (N) and brown (L) (AC) cable and connector from the 2-
pin J1 receptacle on the power supply PCB.
Figure 12-13.
Right Power Cable Assembly Replacement
6.
Remove the blue (N), brown (L), and green (GND) power cables connected to the
power entry module, noting their current positions.
7.
Remove all sets of M6 lug nuts and ground wire cable ring terminals from the
equipotential stud using a deep well 10mm socket wrench.
To replace the right power cable assembly
1.
Properly orient the new right power cable assembly.
2.
Connect one ground wire cable ring terminals to the equipotential stud.
3.
Secure in place with a lug nut, tightening to 12 inch-pounds with a deep well
10mm socket wrench.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......