Performance Tests
Service Manual
10-39
Figure 10-18.
LIGHT Test: BPM 60, SpO
2
90, MOD High, Light High
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately one-inch P-T-P amplitude. Actual amplitude
may vary but references high pulse amplitude/high light indications.
19.
Press MOD.
20.
Verify the MOD decrease LED lights. The pulse amplitude waveform initially decreases in
amplitude and then stabilizes at P-T-P amplitude of approximately half-inch.
a.
Active visual and audio alarms.
b.
Saturation indication between 73 and 77 inclusive.
c.
BPM indication between 57 and 63 inclusive.
d.
Pulse waveform of approximately half-inch P-T-P amplitude. Actual amplitude
may vary but references low pulse amplitude/high light patients.
21.
Press LIGHT LEVEL.
22.
Verify the decrease LED lights. The pulse amplitude waveform initially flatlines and
then stabilizes at the same P-T-P amplitude of approximately half-inch.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......