Operation
6-24
Service Manual
Figure 6-11.
Default Monitoring Screen Layout
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.
To set the default main trend parameters
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select DISPLAY SETTINGS.
4.
Select MAIN SCREEN TREND PARAMETERS.
5.
Select the desired default.
a.
SpO
2
only
b.
Pulse only
c.
SpO
2
and pulse
6.
Select the desired option.
a.
Press SAVE CHANGES to retain the change.
b.
Press CANCEL to leave the default as it was.