Performance Tests
Service Manual
10-13
q.
Slide the bar up to a value of 85, using arrow keys as necessary.
r.
Press SAVE CHANGES.
s.
Repeat steps two through four to view SpO
2
Upper and Lower Alarm Limits.
t.
Select SpO
2
.
u.
Ensure the selected number for SpO
2
Lower Alarm Limit is set at a value of 85.
v.
Slide the bar up until it will not slide any further. Ensure the selected number
for SpO
2
Lower Alarm Limit increases to a value of 99.
w.
Press SAVE CHANGES.
x.
Repeat steps two through four to view SpO
2
Upper and Lower Alarm Limits.
y.
Select SpO
2
.
z.
Ensure the selected number for SpO
2
Lower Alarm Limit is set at a value of 99.
5.
Test Pulse Rate Upper and Lower Alarm Limits.
a.
Select PULSE RATE.
b.
Select the Pulse Rate Upper Alarm Limit, which should be set at a value of 170.
c.
Slide the bar up until it will not slide any further. Ensure the selected number
for Pulse Rate Upper Alarm Limit increases to a value of 250 for the Pulse Rate
Upper Alarm.
d.
Press SAVE CHANGES.
e.
Repeat steps two through four to view SpO
2
Upper and Lower Alarm Limits.
f.
Select PULSE RATE.
g.
Ensure the selected number for Pulse Rate Upper Alarm Limit is set at a value
of 250.
h.
Slide the bar down until it will not slide any further. Ensure the selected
number for Pulse Rate Upper Alarm Limit decreases to a value of 21.
i.
Press SAVE CHANGES.
j.
Repeat steps two through four to view SpO
2
Upper and Lower Alarm Limits.
k.
Select PULSE RATE.
l.
Ensure the selected number for Pulse Rate Upper Alarm Limit is set at a value
of 21.
m.
Press SAVE CHANGES.
n.
Repeat steps two through four to view SpO
2
Upper and Lower Alarm Limits.
o.
Select PULSE RATE.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......