User Interface
Service Manual
6-13
Figure 6-9.
Sample Alarm Limit Violations
6.4.4
Audible Alarm Management
WARNING:
Do not silence or disable audible alarms if patient safety could be
compromised. Do not dim or disable visual alarms if patient safety could be
compromised.
Audible indicators include pitched tones and beeps. Audible alarms vary,
depending on the priority of the alarm. Caregivers may choose to silence
alarms by pressing ALARM SILENCE. For any alarm condition still active for
more than two (2) minutes, the monitoring system will increase the urgency
level of the audible alarm signal by increasing its frequency.
1
Saturation below lower threshold
6
Silence Alarm icon, not active
2
Pulse rate below lower threshold
7
SatSeconds alarm present
3
Patterns of desaturation present
8
Current saturation value, low
4
SPD Alert alarm icon
9
SPD Alert alarm present
5
SatSeconds alarm icon
10
Current pulse rate value, low
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......