Alarms and Error Conditions
Service Manual
11-7
High
System
Failure
Monitor was reset unexpectedly
Unexpected reset. Settings
lost. If problem persists,
remove from service. Refer-
ence
Yes
Low
Extended
Update
Sensor is connected to a patient and has suc-
cessfully detected SpO
2
and pulse rate during
the measurement session, but current condi-
tions are causing the SpO
2
and/or pulse rate
update period(s) to exceed 25 seconds.
Check patient. Reposition
sensor, replace sensor, or
assess alternative Nellcor™
sensor if condition persists.
Yes
Low
System
Failure
The graphical display of SpO
2
or pulse rate
data has not been updated for over 30
seconds
Reset the monitoring
system. If the problem per-
sists, remove from service
and return to a qualified
service technician.
No
Low
Equipment
Failure
Nurse call output system error
A Nurse Call error occurred.
Reference
No
Serial Communications system failure
Serial communication error
occurred. Reference
Yes
Analog output system failure
Analog output error
occurred. Reference
Yes
Cooling fan failure
A cooling fan failure
occurred. Reference
Yes
Low
Over Tem-
perature
Thermal control system is above thermal limit
Internal temperature is
above thermal limit. The
monitor may shut down if
the internal temperature
continues to increase. Refer-
ence
17.
Yes
Low
Clock Set-
tings Lost
Invalid date and time value detected during
start up
Date and time setting lost.
Set the date and time. Refer-
ence
17.
Yes
1. The priority for this alarm can be configured to Low, Medium, or High as an institutional setting. The default priority is Low. Reference
Table 11-2.
Initial Alarm Priority for Errors (Continued)
Priority
Message
Condition
Resolution
Dismiss
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......