![Covidien Nellcor GR101704 Service Manual Download Page 84](http://html1.mh-extra.com/html/covidien/nellcor-gr101704/nellcor-gr101704_service-manual_2673898084.webp)
Operation
6-22
Service Manual
To set the alarm limit threshold defaults
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select ALARM LIMITS.
4.
Select the desired mode: ADULT or NEONATE.
Note:
Pulse rate and SpO
2
thresholds are different for adults and neonates. In addition,
the SPD alarm management parameter is not available for neonates.
5.
Select the desired type of limit.
a.
Select SpO
2
and slide bars up or down to set upper and lower alarm limits.
b.
Select PULSE and slide bars up or down to set upper and lower alarm limits.
c.
Select SATSECONDS and set to 10, 25, 50, or 100 SatSeconds, or OFF.
d.
Select SPD and set the sensitivity to 1, 2, or 3, or OFF.
e.
Select PULSE RATE DELAY to 5 or 10 seconds, or OFF.
6.
Press SAVE CHANGES.
7.
Press EXIT MENU to return to the Service Main Menu.
Sensor Alarm Priorities
Institutional default alarm priorities can be set for the following sensor alarms:
•
Sensor Disconnected
•
Sensor Off
•
Sensor Failure
By default, these sensor alarms are low priority. They can be configured inde-
pendently as high, medium, or low.
To set sensor alarm priorities
1.
Access the service function menu. Reference
To access service function menus
2.
Select SETTINGS.
3.
Select ALARM SYSTEM.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......