Operational Performance Issues
Service Manual
11-17
11.9
Operational Performance Issues
11.10
Hardware Issues
WARNING:
If an monitoring system reports a primary speaker failure, do not use the
monitoring system longer than necessary to ensure patient safety. Contact
Covidien or a local Covidien representative.
Table 11-8.
Common Operational Performance Issues
Problem
Resolution
Monitoring screen appears functional, but is not reg-
istering patient data.
•
Ensure pulse oximetry sensor and interface cable
are both Nellcor™ products.
•
Check the for the loss-of-pulse indicator. If lit,
ensure Nellcor™ sensor is firmly connected. Ref-
erence
Nellcor™ Sensor Performance Consider-
•
Check the monitoring screen for the jagged inter-
ference indicator. If lit, ensure Nellcor™ sensor is
firmly connected and patient remains still.
Patient data appears suspect.
•
Check sensor and interface cable connections.
•
Reset the monitoring system.
Intermittent or corrupt patient data.
Table 11-9.
Common Prompts and Error Messages
Problem
Resolution
Primary speaker failure is detected during
operation.
•
Remove from service immediately.
•
Check cabling and connectivity. Replace if needed.
•
Check the speaker. Replace if needed.
•
Check the Main PCB. Replace if needed.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......