Repair
12-14
Service Manual
Figure 12-9.
Parameter Module Disassembly
10.
Remove and set aside the top cover.
11.
Remove and set aside the four parameter interface PCB assembly T-10 torx
screws.
12.
Remove and set aside the parameter interface PCB assembly on a static-free work
surface.
13.
Remove and set aside failing top cover, faceplate, or bottom enclosure in an alter-
nate location.
To replace a parameter module case component
1.
Select the new parameter module case component: top cover, faceplate, or
bottom enclosure, and place with the still-good components on a static-free work
surface.
2.
Rotate the parameter interface PCB assembly so the NELL1SR PCB is facing down-
ward and both connector sockets are facing the front of the bottom enclosure.
3.
Place on the left-hand bottom enclosure standouts.
4.
Replace and tighten the four T-10 torx screws to a maximum of 10 inch-pounds.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......