Chassis Disassembly and Reassembly
Service Manual
12-11
To remove the parameter module
1.
Gently rotate the monitoring system, placing so the bottom of the chassis is up.
Figure 12-6.
Parameter Module Screw Removal
2.
Remove T-15 screw located between the slots at the rear of the bottom chassis.
3.
Using a flat-blade screwdriver, release the faceplate tabs on either side by pressing
towards the outside and towards the back simultaneously, listening for a click to
indicate tab release.