Basic Preventive Maintenance
Service Manual
12-7
12.4.1
Fuse Removal and Replacement
To remove and replace external fuses
1.
Remove the fuse drawer from the power entry module.
a.
Press down on the center tab.
b.
Pull outward.
2.
Remove both 5 x 20-mm, slow blow, 1.5-amp, 250-volt fuses.
Figure 12-3.
External Fuse Removal
3.
Replace with two (2) new, 5 x 20-mm, slow blow, 1.5-amp, 250-volt fuses.
4.
Reinsert the fuse drawer until the tab snaps in place.
12.4.2
Battery or Battery Access Door Replacement
To remove the battery access door and battery
1.
Using the T-15 torx driver, remove and set aside the torx screw from the battery
cover.