![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 376](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735376.webp)
Instrument Function
R&S
®
SMBV100A
361
Operating Manual 1176.8016.02 ─ 17
Oversampling
Determines the upsampling factor. If the check box is activated, the most sensible
parameter values are selected. The value depends on the coherence check. If the
check box is deactivated, the values can be changed manually.
Remote command:
[:SOURce<hw>]:BB:DM:FILTer:OSAMpling:AUTO
[:SOURce<hw>]:BB:DM:FILTer:OSAMpling
More - Filter
Calls the menu for displaying the filter characteristic of the currently selected filter.
Load User Filter
Calls the menu "Select List File User Filter" for selecting a user-defined filter file with
extansion
*.vaf
on page 350). The button is only available if filter
type "USER" is selected.
The button "File" "Manager" calls the "File" "Manager" menu. The "File" "Manager"
menu is used to display and delete files.
Remote command:
[:SOURce<hw>]:BB:DM:FLISt:SELect
[:SOURce<hw>]:BB:DM:FLISt:CATalog?
[:SOURce<hw>]:BB:DM:FLISt:DELete
Current Filter Response
Displays the filter characteristic of the currently selected filter.
Remote command:
n.a.
Power Ramp
The lower part of the"Custom Digital Modulation" menu is used for setting triggers,
marker and clocks, as well as for power ramp settings.
Power Ramp Control
Calls the power ramp control menu (see
Remote command:
n.a.
Trigger/Marker
Calls the "Trigger/Marker/Clock" dialog, used to select the trigger source, set the time
delay on an external trigger signal and configure the marker output signals.
Baseband Signal - Baseband Block