Remote Control Commands
R&S
®
SMBV100A
590
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Depth>
float
Range:
0 to 100
Increment: See data sheet
*RST:
30
Example:
AM 15PCT
sets the 15% AM modulation depth
Manual operation:
See
[:SOURce<hw>]:AM:EXTernal:COUPling
<Coupling>
Selects the coupling mode for the external amplitude modulation signal.
Parameters:
<Coupling>
AC | DC
AC
Uses only the AC signal component of the modulation signal.
DC
Uses the modulation signal as it is, with AC and DC.
*RST:
AC
Example:
AM:EXT:COUP AC
selects the coupling mode AC for external amplitude modulation.
Manual operation:
See
[:SOURce<hw>]:AM:SENSitivity?
Queries the input sensitivity of the externally applied signal for amplitude modulation.
The sensitivity depends on the set modulation
The returned value reports the sensitivity in
%/V. It is assigned to the voltage value for
full modulation of the input.
Return values:
<Sensitivity>
float
Range:
0 to 100
Example:
AM:DEPT 50
sets a modulation depth of 50 %.
AM:SENS?
queries the input sensitivity at the external modulation input.
Response: 50
since the voltage value for full modulation is 1V, the resulting
sensitivity is precisely 50 %/V.
Usage:
Query only
Manual operation:
See
SOURce Subsystem