Instrument Function
R&S
®
SMBV100A
160
Operating Manual 1176.8016.02 ─ 17
Additionally there are intermediate memories in which the current instrument setting
can be saved and then called up again by just pressing a key. This provides fast
switching between different instrument settings.
Only settings which differ from the preset values and configuration data for the operat-
ing elements (e.g. window positions) are saved. As a result the files remain relatively
small. Furthermore, instrument settings can easily be transferred between different
equipped signal generators since the files contain only relevant information. When loa-
ded, the referenced settings are implemented and all non-referenced parameters are
set to the associated preset values.
If list data is part of the instrument settings, e.g. a list of user correction data, a refer-
ence to this list is saved, not the list itself. The list is reactivated when the associated
settings are loaded, but the list may have been modified or deleted in the meantime or
may not be available on a different instrument. If the list has been modified, the new
entries will be used. An error message appears if an attempt is made to access a non-
existing list or to activate settings which are not supported by the instrument.
●
Network settings and remote settings are not saved and restored.
●
Lists are saved and loaded in the appropriate menus. For example, the user cor-
rection data list is created and saved in the "User Correction" menu.
When loading an instrument setting, it is possible to select whether the current fre-
quency and level setting is to be retained or whether the saved settings are to be acti-
vated. It is possible to delete saved instrument settings. A file can be copied by loading
it with "Recall" and then storing it under a new name.
Settings can be transferred easily between instruments with different equipment
options and/or firmware versions because only the settings which differ from the preset
values are affected. When settings are loaded, only those which are possible on the
instrument are implemented. Error messages indicate the settings which cannot be
implemented.
The saved file is transferred from one instrument to another using the memory stick.
General file management functions such as copying and moving data are available in
the "File Manager" dialog.
5.2.8.1
Save/Recall Menu
The settings available in the File menu "Save/Recall" depend on the operation selected
under "Select Operation".
For more information, see
Chapter 4.7.1, "File Select Dialog"
Select Operation
Selects the file function.