![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 742](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735742.webp)
Remote Control Commands
R&S
®
SMBV100A
727
Operating Manual 1176.8016.02 ─ 17
The ASCII files are stored with the fixed file extensions
*.txt
or
*.csv
in a directory
of the user's choice. The directory applicable to the commands is defined with the com-
mand
MMEMory:CDIR
. A path can also be specified in command
SOUR:CORR:DEXC:AFIL:SEL
, in which case the files are stored or loaded in the
specified directory.
Parameters:
<Filename>
<ascii file name>
Example:
CORR:DEXC:MODE IMP
selects that ASCII files with frequency and level value pairs are
imported and transferred into user correction lists.
CORR:DEXC:AFIL:SEL '/var/user/import_ucor.csv'
selects that ASCII file
ucor.csv
is imported.
CORR:DEXC:SEL '/var/user/import_ucor_imp'
selects that the ASCII file
ucor.csv
is imported into user cor-
rection list
ucor_imp
.
Manual operation:
See
" Select ASCII Source / Destination - User Correction "
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SEParator:COLumn
<Column>
Selects the separator between the frequency and level column of the ASCII table.
Parameters:
<Column>
TABulator | SEMicolon | COMMa | SPACe
*RST:
COMMa
Example:
CORR:DEXC:MODE EXP
selects that the user correction list is exported into an ASCII file.
CORR:DEXC:AFIL:SEL '/var/user/import_ucor.csv'
selects ASCII file ucor.csv as destination for the user correction
list data.
CORR:DEXC:AFIL:SEP:COL TAB
the pairs of frequency and level values are separated by a tabu-
lator.
CORR:DEXC:AFIL:SEP:DEC DOT
selects the decimal separator dot.
CORR:DEXC:SEL '/var/user/import_ucor_imp'
selects that the user correction list ucor_imp is imported into
ASCII file ucor.csv.
Manual operation:
See
" Column Separator- User Correction "
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SEParator:DECimal
<Decimal>
Selects the decimal separator used in the ASCII data between '.' (decimal point) and ','
(comma) with floating-point numerals.
SOURce Subsystem
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...