Remote Control Commands
R&S
®
SMBV100A
660
Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:TDELay
<ExtTimeDelay>
Specifies the trigger delay for external triggering. The value affects all external trigger
signals.
Parameters:
<ExtTimeDelay>
float
Range:
0 to 688
Increment: 250E-12
*RST:
0
Default unit: s
Example:
See
[:SOURce<hw>]:BB:DM:TRIGger:DELay:UNIT
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:RDELay?
Queries the time (in seconds) an external trigger event is delayed for.
Return values:
<ResTimeDelaySec> float
Range:
0 to 688
Increment: 0.25E-9
*RST:
0
Example:
See
[:SOURce<hw>]:BB:DM:TRIGger:DELay:UNIT
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:INHibit
<Inhibit>
Sets the number of symbols by which a restart is to be inhibited following a trigger
event. This command applies only in the case of external triggering.
Parameters:
<Inhibit>
integer
Range:
0 to 67108863
*RST:
0
Default unit: symbol
Example:
BB:DM:TRIG:SOUR EXT
selects an external trigger
BB:DM:TRIG:INH 200
sets a restart inhibit for 200 symbols following a trigger event.
Manual operation:
See
SOURce Subsystem