Instrument Function
R&S
®
SMBV100A
223
Operating Manual 1176.8016.02 ─ 17
The display is disabled when list mode is running, see
Remote command:
Note:
The SCPI command
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
sets the level of the "Level" display, that is the level containing offset.
Limit - RF Level
Sets an upper limit for the RF output power.
You can use it to protect your DUT from damage due to high input power. If you enter
an RF level above this value, the instrument limits the output power to this specified
value, and generates a warning message.
However, the level indication in the status bar is not affected.
Note:
The value is not affected by an instrument preset (PRESET key),
*RST
and the
"Save/Recall" function. It is influenced only by the
and the factory value
is equal to maximum level.
Remote command:
[:SOURce<hw>]:POWer:LIMit[:AMPLitude]
Offset (Level)
Sets the level offset relative to the RF level.
The level offset of a downstream instrument (for example an attenuator or amplifier) is
entered.
The entry does not change the value of the RF level at the RF output. It only changes
the RF level displayed in the display header. The value of the RF level in the header
corresponds to the level at the output of the downstream instrument.
Remote command:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
Attenuator Settings
The attenuator mode is set in the Attenuator Settings section.
Attenuator Mode
Sets the attenuator mode at the RF output.
"Auto"
Standard mode.
The electronically switching attenuator switches with a 5 dB step
width at fixed switching points.
RF Block