Remote Control Commands
R&S
®
SMBV100A
793
Operating Manual 1176.8016.02 ─ 17
Parameters:
<PowCntrlCRange> float
Range:
0 to 50
Increment: 0.01
*RST:
30
Default unit: dB
Example:
POW:SPC:CRAN 15
sets the capture range to +/- 15 dB.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:DELay
<PowCntrlDelay>
Defines a waiting period between the level adjustment of the generator and the next
measurement of the power sensor.
Parameters:
<PowCntrlDelay>
integer
Range:
0 to 1000
*RST:
0
Example:
:POW:SPC:DEL 2 ms
the sensor starts the next reading 2 ms after the level adjust-
ment.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:PEAK
<PowCntrlPeak>
Activates power control by means of the peak power values, provided the power sen-
sor supports this function.
Parameters:
<PowCntrlPeak>
0 | 1 | OFF | ON
*RST:
0
Example:
POW:SPC:PEAK ON
uses the measured peak power for power control.
Manual operation:
See
[:SOURce<hw>]:POWer:SPC:SELect
<PowCntrlSelect>
Defines the currently selected sensor to be used for power control.
Parameters:
<PowCntrlSelect>
SENS1 | SENS2 | SENS3 | SENS4
*RST:
SENS1
SOURce Subsystem