Remote Control Commands
R&S
®
SMBV100A
570
Operating Manual 1176.8016.02 ─ 17
Example:
MMEMory:DATA '/var/user/test.txt',#15hallo
Writes the block data to the file
test.txt
.
The digit
1
indicates a length entry of one digit; the digit
5
indi-
cate a length of the binary data (
hallo
) in bytes.
MMEMory:DATA? '/var/user/test.txt'
Sends the data of the file
test.txt
from the instrument to the
controller in the form of a binary block.
Response: #15hallo
Usage:
SCPI confirmed
:MMEMory:DATA:UNPRotected
<Msus>, <Data>
The
setting
command sends the I/Q data and the marker data to the file defined with
the
<Msus>
parameter. The required tags are created automatically so that the file
content follows the waveform file format.
The
query
reads out the I/Q data part or the marker information of the specified file.
Waveform files can also be created with the command
. In this case, the
complete content
of the waveform file
must be specified, i.e. not only the I/Q or marker data but all required tags.
Parameters:
<Msus>
<Identifier>:<file name>
Mass Storage Unit Specifier
A string that specifies whether I/Q data (
NVWFM
) or marker data
(
NVMKR
) is transferred and the file name the data is stored in.
If the file with the specified name does not exist, a file is created.
Any existing content in the file is
overwritten
.
Complete file path and file extension may also be specified. If
omitted, files are stored in the default directory and the exten-
sion
*.wv
is assigned to the file name.