Remote Control Commands
R&S
®
SMBV100A
787
Operating Manual 1176.8016.02 ─ 17
Manual operation:
See
[:SOURce<hw>]:POWer:ALC[:STATe]
<State>
Activates/deactivates automatic level control.
Parameters:
<State>
0 | OFF | AUTO | 1 | ON | OFFTable | OFFHighAccuracy
ON
Internal level control is permanently activated.
OFF|ONSample
Internal level control is deactivated.
AUTO
Internal level control is activated/deactivated automatically
depending on the operating state.
ONTable
Internal level control is performed according to the ALC table.
OFFHighAccuracy
The ALC mode High Accuracy ensures improved level accuracy.
Note:
This mode increases the frequency and level setting time
up to 20 ms!
*RST:
AUTO
Manual operation:
See
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
<Mode>
Selects the attenuator mode, when the RF signal is switched off.
Parameters:
<Mode>
UNCHanged | FATTenuation
UNCHanged
Freezes the setting of the attenuator when RF is switched off.
The attenuator is only activated when RF is switched on.
This setting recommended if a constant VSWR (
V
oltage
S
tand-
ing
W
ave
R
atio) is required.
FATTenuation
Sets attenuation to maximum when the RF signal is switched off.
This setting is recommended for applications that require a high
level of noise suppression.
*RST:
n.a. (factory preset: FATTenuation)
Example:
SOUR:POW:ATT:RFOF:MODE FATT
sets the RF OFF attenuator to maximum.
Manual operation:
See
SOURce Subsystem