Instrument Function
R&S
®
SMBV100A
221
Operating Manual 1176.8016.02 ─ 17
In addidion, when the self test is running, the displayed level in the header is 10 dBm,
but the actual signal level at the RF output is -50 dBm.
dBm, dBµV, mV and µV can be used as the level units. The four unit keys are labeled
with these units.
Note:
The SCPI command sets the level of the "Level" display, i.e. an entered level off-
set is considered in the level value.
Remote command:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
5.5.6.2
RF Level Dialog
Access:
► Select "RF/A Mod > config... > RF Level > Level/Attenuator".
The offset-free level, level offset and level limit are set in the top section of the dia-
log. The attenuator mode is set in the "Attenuator Settings" section.
RF Block