![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 850](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735850.webp)
Remote Control Commands
R&S
®
SMBV100A
835
Operating Manual 1176.8016.02 ─ 17
Example:
SYST:IDEN USER
selects an user-defined identification
SYST:IRES "Test Device"
defines the identification string 'test device'
*IDN?
Response: 'test device'
Manual operation:
See
:SYSTem:ORESponse
<OResponse>
Defines the user defined response string for *OPT.
Note
: While working in a emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command
SYST:ORES
will be discarded.
Parameters:
<OResponse>
string
Example:
SYST:IDEN USER
selects an user-defined identification
SYST:ORES "Test Option"
defines the OPT string 'test option'
*OPT?
Response: 'test option'
Manual operation:
See
:SYSTem:LANGuage
<Language>
Sets the remote control command set.
The instrument can also be remote controlled via the command set of several other
generators, for example HP generator. See the Application Note
at the down-
load area of the product site on the Internet.
Note
: While working in a emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command
SYSTem:LANGuage
will be discarded.
The return to the SCPI command set of the R&S
SMBV can only be performed by
using the appropriate command of the selected command set. For example, the HP
command
EX
returns to the instrument-specific GPIB command set (selection
SYST:LANG 'HPxxxx'
).
Parameters:
<Language>
string
Example:
SYSTem:LANGuage "SCPI"
sets the SCPI command set.
Manual operation:
See
SYSTem Subsystem