![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 274](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735274.webp)
Instrument Function
R&S
®
SMBV100A
259
Operating Manual 1176.8016.02 ─ 17
Note:
The installed hardware determines the available settings. Use the
dialog to check the hardware the instrument is equipped with.
For information on the required hardware revision, refer to the release notes.
Remote command:
ɸ
M Sensitivity
Displays the input sensitivity of the externally applied modulation signal at the EXT
MOD input in RAD/V.
The modulation deviation entered with
"
ɸ
M Deviation" is achieved with 1
Volt (=U
peak
) of the input signal.
Note:
The input voltage must not exceed 1.1 V
p
otherwise modulation distortions
occur.
Remote command:
Mod External Coupling
Selects the coupling mode ("AC" or "DC") for the external modulation signal.
Note:
Coupling for external feed via input EXT MOD can be set independently for all
modulations using the external modulation signal.
"AC"
Disconnects the DC voltage component and uses only the AC com-
ponent of the modulation signal.
"DC"
Uses the modulation signal with both components, AC and DC.
Remote command:
[:SOURce<hw>]:PM:EXTernal:COUPling
Ext. Impedance
(Source "External" only)
Sets the impedance for the external modulation signal, applied at the EXT MOD con-
nector.
You can select 600 Ohm or high (>100 kOhm).
This setting affects all analog modulations which use the external modulation signal.
Remote command:
[:SOURce<hw>]:INPut:MODext:IMPedance
5.5.12.5
Pulse Modulation (PM)
External and internal Pulse modulation requires options R&S
SMBV-K22, Pulse Modu-
lator and R&S
SMBV-K23, Pulse Generator.
RF Block