![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 624](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735624.webp)
Remote Control Commands
R&S
®
SMBV100A
609
Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:ARBitrary:TRIGger:DELay:UNIT
<TrigDelUnit>
Determines the units the trigger delay is expressed in.
Parameters:
<TrigDelUnit>
SAMPle | TIME
*RST:
SAMPle
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:DELay
<Delay>
The command specifies the trigger delay (expressed as a number of samples) for
external triggering.
Parameters:
<Delay>
float
Range:
0 to depends on other values
Increment: 0.01
*RST:
0
Default unit: sample
Example:
BB:ARB:TRIG:SOUR EXT
selects an external trigger
BB:ARB:TRIG:DEL 200
sets a delay of 200 samples for the trigger.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:TDELay
<ExtTimeDelay>
Specifies the trigger delay for external triggering. The value affects all external trigger
signals.
Parameters:
<ExtTimeDelay>
float
Range:
0 to 688
Increment: 1E-9
*RST:
0
Default unit: s
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:RDELay?
Queries the time (in seconds) an external trigger event is delayed for.
Return values:
<ResTimeDelaySec> float
Range:
0 to 688
Increment: 0.25E-9
*RST:
0
SOURce Subsystem