![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735076.webp)
Preparing for Use
R&S
®
SMBV100A
61
Operating Manual 1176.8016.02 ─ 17
To trace and display messages:
1. In the navigation pane, select "Diagnostics" > "SCPI Remote Trace".
2. In the toolbar bar of the "SCPI Remote Trace" page, select "live mode" > "on" and
"logging" > "on".
"live mode > on" displays all commands and responses, and "logging > on" also
traces messages.
3. If you now control the R&S
SMBV with SCPI commands, using an appropriate tool,
the SCPI remote trace records the information sent and received.
The function records all sent commands, received responses and messages, and
stores them in an internal database. If "live mode" is disabled, you can display the
recent traces upon request, using the "refresh" button. You can also store the log in
a file.
Note:
The diagnostics functionality is extended in later releases, e.g. to download
or upload SCPI command files from / to the instrument.
Toolbars
The toolbar at the top of the dialog provides basic settings and functions.
●
"Live mode" / "logging": If logging is switched on, messages are traced. They are
stored in an internal database and can be displayed upon request, using the
LXI Configuration