Remote Control Commands
R&S
®
SMBV100A
640
Operating Manual 1176.8016.02 ─ 17
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CATalog?
Queries the available configuration files in the specified default directory. The configu-
ration files are used to create multi segment waveform files.
Return values:
<Catalog>
string
Example:
MMEM:CDIR '/var/user/waveform'
sets the default directory.
BB:ARB:WSEG:CONF:CAT?
reads out all the configuration files in the default directory.
Response: mulit1, multi2
the directory contains the configuration files
multi1.inf_mswv
and
multi2.inf_mswv
.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CLOCk
<Clock>
This command defines the clock rate used for multi segment waveform output in case
of Clock Mode "User" (
:BB:ARB:WSEG:CONF:CLOCk:MODE USER
).
Parameters:
<Clock>
float
Increment: 1E-3
*RST:
max SampleRate
Example:
BB:ARB:WSEG:CONF:CLOC:MODE USER
selects Clock Mode User.
BB:ARB:WSEG:CONF:CLOC 50MHz
defines a clock rate of 50 MHz.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:CLOCk:MODE
<Mode>
Selects the clock rate mode for the multi segment waveform.
SOURce Subsystem