Instrument Function
R&S
®
SMBV100A
317
Operating Manual 1176.8016.02 ─ 17
Optimize I/Q Signals for RF Output
Activates/deactivates optimization of level accuracy and imbalance for RF Output.
Level accuracy and imbalance can be either optimized for RF output (active) or for dif-
ferential outputs (not active).
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:BIAS:OPTimization[:RF]:STATe
I/Q Output Type
Selects the type of output. The analog outputs can be operated in "Single Ended"
mode or "Differential" (non-inverted) mode. The menu changes depending on the
selected output mode
"Differential"
Differential output at I
OUT and I
OUT
BAR or Q
OUT and
Q
OUT
BAR.
"Single Ended" Single-ended output at I OUT and Q OUT.
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE
Mode
Selects the mode for setting the outputs - independent or together. The menu changes
depending on the selection here.
"Fixed"
The settings for the I/Q signal components are fixed and cannot be
changed.
●
"Level Vp (EMF)" (Single ended) = 1.0 V
●
"Level Vp (EMF)" (Differential) = 2.0 V
"Variable"
The settings for the I/Q signal components are not fixed and can be
modified.
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block