![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 293](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735293.webp)
Instrument Function
R&S
®
SMBV100A
278
Operating Manual 1176.8016.02 ─ 17
"Single"
Generates a single sweep cycle after a trigger event.
The sweep steps within the cycle are performed automatically, con-
trolled by the dwell time. If one cycle is completed, the instrument
waits for the next trigger event.
To trigger the sweep, use "Execute Single Sweep" button, or the cor-
responding remote control commands, for example
*TRG
.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC
"Step"
Generates the sweep signal step-by-step, manually triggered.
To perform the sweep steps, enter the frequency value under
. You can directly enter the value, but also
use the UP and DOWN navigation keys or the ROTARY KNOB.
You can determine the step width below in the entry field "Step Lin" or
"Step Log", see
Step Lin/Log - Frequency Sweep
If a step is out of the sweep range ("Start Freq" or "Stop Freq"), it is
ignored.
Note:
To step through the sweep frequencies in remote control
mode, use the
FREQ:MAN
command with the
UP
or
DOWN
parameter.
Example:
SOUR:FREQ:CENT 300MHz
SOUR:FREQ:SPAN 400MHz
SOUR:SWE:FREQ:SPAC LIN
SOUR:SWE:FREQ:STEP:LIN 100MHz
SOUR:FREQ:MODE MAN
TRIG:FSW:SOUR SING
set sweep mode "Step".
SOUR:FREQ:MODE SWE
activate sweep mode, the frequency is set to "Start Freq".
SOUR:FREQ:MAN UP
set the frequency to the next higher sweep frequency.
SOUR:FREQ:MAN DOWN
set the frequency to the next lower sweep frequency.