![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 704](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735704.webp)
Remote Control Commands
R&S
®
SMBV100A
689
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Source>
SOFTware | MARKer
SOFTware
Recording of signals is started automatically in specified inter-
vals. This asynchronous method is appropriate when a repre-
sentative display of the complete signal is desired, as recording
starts in a random time reference to the signal.
MARKer
The starting point for recording signals is determined by "Marker
1" in the menu of the active standard.
This synchronous method is appropriate when specific signal
segments are to be shown in greater detail. With periodic sig-
nals, a static image is obtained by selecting a period length of
"Marker 1" that is equal to the signal period.
The displayed signal segment can be shifted as needed by
entering a "Marker Delay" in the "Trigger/Marker" menu of the
active standard.
*RST:
SOFTware
Example:
BB:GRAP:TRIG:SOUR MARK
sets the trigger source.
Manual operation:
See
7.16.3.5
SOURce:BB:IMPairment Subsystem
This subsystem contains the commands for the digital I/Q impairments.
Included in this section is also the description of command
SOURce:BB:IQOutput:SOURce
that determined the source for the I/Q output con-
nectors.
[:SOURce<hw>]:BB:IMPairment:DELay
............................................................................689
[:SOURce<hw>]:BB:IMPairment:IQRatio[:MAGNitude]
.......................................................690
[:SOURce<hw>]:BB:IMPairment:LEAKage:I
......................................................................690
[:SOURce<hw>]:BB:IMPairment:LEAKage:Q
.................................................................... 690
[:SOURce<hw>]:BB:IMPairment:QUADrature[:ANGLe]
...................................................... 691
[:SOURce<hw>]:BB:IMPairment:SKEW
............................................................................691
[:SOURce<hw>]:BB:IMPairment:STATe
........................................................................... 692
[:SOURce<hw>]:BB:IMPairment:OPTimization:MODE
....................................................... 692
[:SOURce<hw>]:BB:IMPairment:OPTimization:STATe
.......................................................692
[:SOURce<hw>]:BB:IMPairment:DELay
<Delay>
Sets the time delay of both I and Q vectors relative to the selected trigger and marker
or relative to the other instrument(s) working in the master-slave synchronous mode
(see section
Chapter 5.9.2.4, "Synchronous Signal Generation"
A positive value means that the I and Q vectors delay relative to the marker/trigger or
to the other instrument and vice versa.
SOURce Subsystem
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...