Example:
:STAT:OPER:NTR 0
a transition from 1 to 0 in the condition part of the Status:Opera-
tion register does not cause an entry to be made in the EVENt
part.
:STATus:OPERation:PTRansition
<Ptransition>
Sets the bits of the PTRansition part of the STATus:OPERation register. If a bit is set,
a transition from 0 to 1 in the condition part causes an entry to be made in the EVENt
part of the register. A new event in the hardware is thus registered, for example the
start of an adjustment.
Parameters:
<Ptransition>
string
Example:
:STAT:OPER:PTR 32767
all transitions from 0 to 1 in the condition part of the Status:Oper-
ation register cause an entry to be made in the EVENt part.
:STATus:PRESet
<Preset>
Resets the status registers. All PTRansition parts are set to FFFFh (32767), i.e. all
transitions from 0 to 1 are detected. All NTRansition parts are set to 0, i.e. a transition
from 1 to 0 in a CONDition bit is not detected. The ENABle parts of STATus:OPERa-
tion and STATus:QUEStionable are set to 0, i.e. all events in these registers are not
passed on.
Parameters:
<Preset>
string
Example:
STAT:PRES
resets the status registers.
:STATus:QUEStionable:CONDition
<Condition>
Queries the content of the CONDition part of the STATus:QUEStionable register. This
part contains information on the action currently being performed in the instrument.
The content is not deleted after being read out since it indicates the current hardware
status.
Parameters:
<Condition>
string
Example:
:STATus:QUEStionable:CONDition?
queries the Status:Questionable:Condition register.
:STATus:QUEStionable:ENABle
<Enable>
Sets the bits of the ENABle part of the STATus:QUEStionable register. The enable part
determines which events of the STATus:EVENt part are enabled for the summary bit in
the status byte. These events can be used for a service request.