Instrument Function
R&S
®
SMBV100A
405
Operating Manual 1176.8016.02 ─ 17
Generate Signal HD (Sine Test Signal)
Generates a signal and saves it to a file.
The File Select window opens automatically and the signal can be stored as a wave-
form file.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate:NAMed
Generate Signal RAM (Sine Test Signal)
Generates a signal and uses it as output straight away.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate
Rect Test Signal
To access the "Sine Test Signals" dialog, in the "ARB MOD" dialog set the parameter
"Create Test Signal" to Rect and press "Create Test Signal".
The dialog can be used to configure a rectangular test signal. A rectangular test signal
with a duty factor of 0.5 is created. Amplitude and offset are adjustable. Both paths, I
and Q, get the same signal.
Frequency (Rect Test Signal)
Enters the frequency of the test signal.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:FREQuency
Samples per Period (Rect Test Signal)
Enters the number of sample values required for the rectangular signal per period.
The resulting clock rate must not exceed the maximum ARB clock rate (see data
sheet). The number of sample values is automatically restricted by reference to the set
frequency.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:SAMPles
Amplitude (Rect Test Signal)
Enters the digital amplitude of the rectangular wave.
The abbreviation FS means full scale.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:AMPLitude
Baseband Signal - Baseband Block