Parameters:
<Phase>
float
Range:
-180 to 180
Increment: 0.01
*RST:
90
Default unit: DEG
Example:
BB:ARB:TSIG:SINE:PHAS 90
sets a phase offset of 90 degrees.
Manual operation:
See
"Phase Offset Q (Sine Test Signal)"
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:SAMPles
<Samples>
The command sets the sample rate for the sine signal in samples per period.
The resulting clock rate must not exceed the maximum ARB clock rate of 150 or 90
MHz depending on the installed options.
The maximum value is automatically restricted by reference to the set frequency and
has to fulfill the rule Frequency * Samples <= ARB clock rate.
Parameters:
<Samples>
integer
Range:
4 to 1000
*RST:
100
Default unit: 100 samples per period
Example:
BB:ARB:TSIG:SINE:SAMP 100
sets a sample rate of 100 samples per period.
Manual operation:
See
"Samples per Period (Sine Test Signal)"
SOURce:BB:ARB:MCAR Subsystem
The
MCARrrier
subsystem contains the commands for setting the Multi Carrier Wave-
form Generator.
Chapter 7.4, "Waveform and List Format"
waveform format and how it is used to transmit waveforms via the IEC bus.
CARRier<ch>
The numerical suffix under
CARRier
distinguish between the carriers. The value range
is 0 .. 31.
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:COUNt
.................................................619
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:MODE
................................................. 619
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:SPACing
.............................................. 620
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:CONFlict?
..................................... 620
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:DELay
.......................................... 621
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:FILE
............................................. 621
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:FREQuency
.................................. 621
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier<ch>:PHASe
......................................... 622
SOURce Subsystem
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...