![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 759](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735759.webp)
Remote Control Commands
R&S
®
SMBV100A
744
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Impedance>
G50 | G1K
G1K
1 kOhm to ground
G50
50 Ohm to ground
*RST:
G1K
Example:
INP:TRIG:IMP G50
all trigger and control signal inputs are set to 50 ohm to ground.
This setting is recommended in the case of high clock rates.
Manual operation:
See
[:SOURce]:INPut:TRIGger:LEVel
<Level>
Sets the high/low threshold of the trigger and control signal inputs in the baseband sec-
tion. In the case of positive polarity, this threshold determines the point as of which a
signal is high (active) or low (inactive).
The command affects:
●
TRIGGER inputs
●
CLOCK inputs
Parameters:
<Level>
float
Range:
0.01 to 2
Increment: 0.01
*RST:
1
Default unit: V
Example:
INP:TRIG:LEV 1 V
a high/low threshold of 1 volt is set at all trigger and control sig-
nal inputs of the baseband section. In the case of positive polar-
ity, the signal is high (active) for a signal voltage of 1 volt and
higher.
Manual operation:
See
"Threshold Trigger/Clock Input"
[:SOURce]:INPut:TRIGger:SLOPe
<Slope>
Sets the polarity of the active slope of an externally applied trigger signal at the trigger
input (BNC connector at the rear of the instrument).
Parameters:
<Slope>
NEGative | POSitive
*RST:
POSitive
Example:
INP:TRIG:SLOP NEG
Activates the falling slope of the external trigger signal at the
trigger input.
SOURce Subsystem