Remote Control Commands
R&S
®
SMBV100A
791
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Manual>
float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model
The values are given in the data sheet.
Increment: 0.01
*RST:
-30
Example:
POW:SWE:MODE MAN
sets the Step sweep mode for RF output A.
POW:MAN -5 dBm
sets an RF level of -5 dBm for the next setting in the Step sweep
mode for RF output A.
POW:MODE SWE
sets the Level Sweep mode for RF output A.
POW:MAN -5.5 dBm
triggers the next sweep step with a level of -5.5 dBm.
Manual operation:
See
" Current Level - Level Sweep "
[:SOURce<hw>]:POWer:MODE
<Mode>
Sets the instrument operating mode and therefore also the commands used to set the
output level.
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW
and
FIXed
are synonyms. To set the output level value, use
the command
SWEep
Operates in power sweep mode.
Set the range and current level with the commands
.
*RST:
CW
Example:
POW:MODE SWEep
selects the SWEep mode using the
POW:STAR; POW:STOP; POW:MAN
settings.
Manual operation:
See
SOURce Subsystem