Instrument Function
R&S
®
SMBV100A
283
Operating Manual 1176.8016.02 ─ 17
"Negative"
activates the falling edge of the trigger signal.
Remote command:
5.5.14.3
RF Level Sweep
To open the "Level Sweep" menu, select "RF/A Mod > Configure > Level Sweep" or
use the MENU key under "RF/A Mod".
RF Level Sweep Settings
The "RF Level Sweep" dialog is used to activate and configure a sweep for the RF
level.
In the top section, the RF level sweep is activated and the sweep mode is selected.
The buttons are used to reset the level sweep (all sweep modes) or to execute the
level sweep ("Single" mode).
The sweep range, sweep spacing and dwell time are set in the bottom section.
State - Level Sweep
Activates Level Sweep mode.
Note:
Activating a sweep mode automatically deactivates other sweeps and the list mode.
Remote command:
Mode - Level Sweep
Selects the level sweep instrument operating mode and the sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.