Remote Control Commands
R&S
®
SMBV100A
543
Operating Manual 1176.8016.02 ─ 17
:BLER:STOP
:BERT:STOP
Stops an ongoing measurement.
These commands automatically set
BERT:STATe
or
BLER:STATe
to OFF
.
Example:
BERT:STOP
stops the measurement. To start a new measurement, the BER
measurement must be switched on again by
BERT:STATe ON
.
Usage:
Event
Manual operation:
See
:BLER:UNIT
<Unit>
:BERT:UNIT
<Unit>
Sets the unit for the error rate display on the screen.
BERT|BLER:RES
always speci-
fies the error rate as the quotient of "Number of errors" and "Number of data items",
unaffected by this command.
Parameters:
<Unit>
ENGineering | SCIentific | PCT | PPM
ENGineering
The error rate is output in exponential notation with the exponent
–3 or –6.
SCIentific
The error rate is output in scientific notation, i.e. standardized to
one place to the left of the decimal.
PCT
The error rate is output in percent.
PPM
The error rate is output in parts per million.
*RST:
ENGineering
Example:
BERT:UNIT PPM
selects the unit "ppm" for the display of the error rate.
Manual operation:
See
[:SOURce]:INPut:BERT:IMPedance
<Impedance>
Sets the impedance of the BERT inputs CLOCK, DATA, DATA ENABLE and
RESTART.
Parameters:
<Impedance>
G1K | G50 | G10K
*RST:
G1K
BERT and BLER Subsystems