Remote Control Commands
R&S
®
SMBV100A
786
Operating Manual 1176.8016.02 ─ 17
..................................................................................... 790
.....................................................................................790
.......................................................................................791
........................................................................................ 792
......................................................................................792
[:SOURce<hw>]:POWer:SPC:CRANge
............................................................................ 792
............................................................................... 793
................................................................................ 793
[:SOURce<hw>]:POWer:SPC:SELect
.............................................................................. 793
...............................................................................794
[:SOURce<hw>]:POWer:SPC:TARGet
............................................................................. 794
.......................................................................................794
[:SOURce<hw>]:POWer:STEP[:INCRement]
.................................................................... 795
............................................................................. 795
........................................................................................796
[:SOURce<hw>]:POWer:ALC:OMODe
<OffMode>
The command sets the level control mode which becomes active when automatic level
control is deactivated (ALC Off).
Parameters:
<OffMode>
SHOLd | TABLe | HIGHaccuracy
SHOLd
Level control is activated briefly if the level or frequency changes
("ALC Off Sample & Hold").
HIGHaccuracy
The ALC mode High Accuracy ensures improved level accuracy.
Note:
This mode increases the frequency and level setting time
up to 20 ms!
TABLe
Internal level control is performed according to the ALC table.
*RST:
SHOLd
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:OMOD SHOL
level control is briefly activated if the frequency or level changes.
Manual operation:
See
[:SOURce<hw>]:POWer:ALC:SONCe
Temporarily activates level control for correction purposes.
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:SONC
level control is performed once only.
Usage:
Event
SOURce Subsystem