Remote Control Commands
R&S
®
SMBV100A
547
Operating Manual 1176.8016.02 ─ 17
Example:
CAL:IQM:FULL?
starts adjustments for the I/Q modulator across the entire fre-
quency range.
Response: "0"
adjustment has been performed successfully
Usage:
Query only
Manual operation:
See
"Adjust I/Q Modulator Full Range"
:CALibration<hw>:IQModulator:LOCal?
Starts adjustment of the I/Q modulator for the currently set frequency, I/Q swap and
baseband gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q
imbalance and quadrature.
Return values:
<Local>
0 | 1
*RST:
0
Example:
CAL:IQM:LOC?
starts adjustment for the I/Q modulator for the currently set fre-
quency.
Response:
0
adjustment has been performed successfully
Usage:
Query only
Manual operation:
See
"Adjust I/Q Modulator Current Frequency"
:CALibration<hw>:LEVel:ALCTable[:MEASure]?
Performs ALC adjustments according to the predefined values in the ALC table.
Return values:
<Measure>
0 | 1 | OFF | ON
Example:
CAL:LEV:ALCT:MEAS?
starts adjustments according to the ALC table.
Respone: "0"
adjustment has been performed successfully.
Usage:
Query only
Manual operation:
See
:CALibration<hw>:LEVel[:MEASure]?
[<Force>]
Starts all adjustments which affect the level.
Query parameters:
<Force>
string
*RST:
force
CALibration Subsystem