Operating Manual 1176.8016.02 ─ 17
SENSOR
Connector for R&S NRP-Zxx sensors.
With the aid of the "User Correction" function, a table with correction values for exter-
nal test assemblies can be automatically determined, e.g. for compensating the fre-
quency response of cables.
The power sensors are connected by inserting the male connector. To disconnect hold
the connector by its sleeve. Pulling on the sensor cable will not release the sensor con-
nector.
LO IN, LO OUT
Local oscillator input and output:
●
LO IN: Input for external LO signals
●
LO OUT: Output for internal LO signals.
IEC 625/IEEE 488
IEC-bus (IEEE 488) interface for remote control of the instrument.
See also
Chapter A.1, "GPIB Bus Interface"
on page 861 and chapter "Remote Con-
trol Basics" in the Operating Manual.
Note:
In order to avoid electromagnetic interference (EMI) caused by open lines,
always terminate any connected IEC-bus cable with an instrument or a controller.
SIGNAL VALID
Output of valid signal. This signal marks the valid signal times (valid level and fre-
quency indication). The signal is generated automatically.
INSTR TRIG
Input for external trigger of sweeps and list mode.
PULSE VIDEO
Output of internal pulse generator signal or external pulse signal fed in via the PULSE
EXT connector (video signal).
PULSE EXT
Input of external pulse signal or input of external trigger/gate signal for internal pulse
generator.
REF OUT
Output of internal reference signal.