![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 639](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735639.webp)
Remote Control Commands
R&S
®
SMBV100A
624
Operating Manual 1176.8016.02 ─ 17
A Target Crest Factor above the crest factor of the unclipped multicarrier signal has no
effect.
Parameters:
<CFactor>
float
Range:
-50 to 50
Increment: 0.01
*RST:
50
Default unit: dB
Example:
BB:ARB:MCAR:CLIP:STAT ON
enables clipping.
BB:ARB:MCAR:CLIP:CFAC 37 dB
sets the target crest factor.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CLIPping:CUToff
<Cutoff>
Sets the cut off frequency of the final lowpass filter, if baseband clipping is enabled
(
BB:ARB:MCAR:CLIP:STAT ON
).
When the cut off frequency is set as half of the output sample rate, a final lowpass filter
improves the spectrum of the clipped multicarrier signal, but may also increase the
resulting crest factor.
Parameters:
<Cutoff>
float
Range:
0 to 50E6
Increment: 0.01
*RST:
50E6
Default unit: MHz
Example:
BB:ARB:MCAR:CLIP:STAT ON
enables clipping
BB:ARB:MCAR:CLIP:CUT 50 MHz
sets the cut off frequency of the filter.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CLIPping[:STATe]
<State>
Switches baseband clipping on and off.
Clipping reduces the peak power of the resulting multi carrier signal according to the
value set with the command
BB:ARB:MCAR:CLIP:CFAC
.
SOURce Subsystem