Remote Control Commands
R&S
®
SMBV100A
778
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Power>
<Power#1>{, <Power#2>, ...} | block data
The data can be given either as a list of numbers (list can be of
any length and list entries must be separated by commas) or as
binary block data. When block data is transferred, 8 bytes are
always interpreted as a floating-point number with double accu-
racy (see
Range:
Minimum level to Maximum level
Default unit: dBm
Example:
LIST:SEL '/var/user/list3'
selects
list3
for editing. The R&S
SMBV generates a new file
automatically, if it does not exist yet.
LIST:POW 0dBm, 2dBm, 2dBm, 3dBm,..
specifies the level values in
list3
. The number of level values
must correspond to the number of frequency values. The previ-
ous data is overwritten.
Manual operation:
See
" Edit List Mode Data... - List Mode "
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the length (in points) of the LEVel part of the selected list.
Return values:
<Points>
integer
Range:
0 to INT_MAX
*RST:
0
Example:
LIST:SEL '/var/user/list3'
selects
list3
for editing. The R&S
SMBV generates a new file
automatically, if it does not exist yet.
LIST:POW:POIN?
queries the number of levels in the list file
Response:
327
Usage:
Query only
[:SOURce<hw>]:LIST:RESet
Resets the list to the starting point.
Example:
LIST:RES
resets the list to the starting point.
Usage:
Event
Manual operation:
See
SOURce Subsystem