Remote Control Commands
R&S
®
SMBV100A
798
Operating Manual 1176.8016.02 ─ 17
SOURce:PULM:STATe 1
OUTPut1:STATe 1
........................................................................................ 798
[:SOURce<hw>]:PULM:DOUBle:DELay
............................................................................798
[:SOURce<hw>]:PULM:DOUBle:STATe
........................................................................... 799
[:SOURce<hw>]:PULM:DOUBle:WIDTh
........................................................................... 799
........................................................................................ 799
[:SOURce<hw>]:PULM:OUTPut:SYNC[:STATe]
................................................................799
....................................................................................... 800
..................................................................................... 800
..................................................................................... 801
........................................................................................ 801
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity
................................................ 801
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance
...................................................... 802
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe
............................................................802
[:SOURce<hw>]:PULM:TRIGger:MODE
........................................................................... 802
........................................................................................ 803
[:SOURce<hw>]:PULM:DELay
<Delay>
Sets the pulse delay.
Parameters:
<Delay>
float
Range:
0 to 100 s
Increment: 10 ns
*RST:
10 ns
Example:
PULM:DEL 13 us
13 us elapse after a trigger before the first pulse is generated.
Options:
R&S R&S
SMBV-K23 (Pulse Generator)
Manual operation:
See
" Pulse Delay - Pulse Generator "
[:SOURce<hw>]:PULM:DOUBle:DELay
<Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay>
float
Range:
10 ns to 100 s
Increment: 5 ns
*RST:
3 us
Example:
PULM:DOUB:DEL 22 us
22 us elapse between the beginning of the first pulse and the
beginning of the second pulse in double-pulse mode.
Options:
R&S
SMBV-K23 (Pulse Generator)
Manual operation:
See
" Double Pulse Delay - Pulse Generator "
SOURce Subsystem
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...