Instrument Function
R&S
®
SMBV100A
238
Operating Manual 1176.8016.02 ─ 17
Remote command:
n.a.
Range
Sets the range for filling the table.
Remote command:
n.a.
Select column to fill
Selects either the frequency or the level column to be filled with the value defined
below.
Remote command:
n.a.
Start value
Sets the start value for the frequency or the level entries.
Remote command:
n.a.
End value
Displays the end value for the frequency or the level entries.
Remote command:
n.a.
Increment value
Sets the increment for the frequency or the level entries.
Remote command:
n.a.
Fill
Fills the selected column in the set range with values, starting with the start value and
using the set increment.
Remote command:
n.a.
5.5.9.3
Filling the Correction List with Power Sensor Measurement Data
The level correction values for the user correction list can be acquired by means of
R&S NRP power sensors. The R&S NRP sensors are connected to either the
SENSOR connector or to one of the USB interfaces. Configuration of the connection is
performed in the "Power Sensor" menu (see
Chapter 5.5.11.1, "NRP-Z Power Viewer"
on page 240). The filling of the user correction list with measurement data is per-
formed in the ucor list editor (see
" Edit User Cor. Data - User Correction "
In the editor, the frequencies for which the correction values are to be acquired are
entered in the frequency column (either manually or by means of the "Fill..." menu).
RF Block