![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 844](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735844.webp)
Remote Control Commands
R&S
®
SMBV100A
829
Operating Manual 1176.8016.02 ─ 17
Note:
it is recommended that you do not change the host name in order to avoid prob-
lems with the networdk connection. However, if you change the host name be sure to
use an unique name.
The host name is a protected parameter, To change it, first disable protection level 1
with command
Parameters:
<Hostname>
string
Example:
SYSTem:PROTect1:STATe OFF,123456
SYSTem:COMMunicate:NETWork:HOSTname 'SIGGEN'
sets the individual computer name of the R&S
SMBV.
Manual operation:
See
:SYSTem:COMMunicate:NETWork[:COMMon]:WORKgroup
<Workgroup>
Sets the individual workgroup name of the instrument.
Parameters:
<Workgroup>
string
Example:
SYSTem:COMMunicate:NETWork:COMMon:WORKgroup
'TEST_09'
sets the workgroup name
Manual operation:
See
:SYSTem:COMMunicate:NETWork[:IPADdress]:DNS
<DNS>
Determines the net DNS server to resolve the name.
Parameters:
<DNS>
string
Example:
SYST:COMM:NETW:IPAD:DNS 123.456.0.1
Manual operation:
See
:SYSTem:COMMunicate:NETWork:IPADdress:MODE
<Mode>
Selects manual or automatic setting of the IP address.
Parameters:
<Mode>
AUTO | STATic
*RST:
n.a. (factory preset: AUTO)
Example:
SYSTem:COMMunicate:NETWork:IPADdress:MODE AUTO
The IP address is assigned automatically (DHCP)
Manual operation:
See
SYSTem Subsystem