Instrument Function
R&S
®
SMBV100A
289
Operating Manual 1176.8016.02 ─ 17
provided e.g. by the Microsoft
®
Excel program. The separators for table columns and
for decimal floating-point numerals can be set. In addition, internally created List data
can be exported into ASCII files using the export function.
The necessary hardware settings are calculated the first time a list is processed. With
long dwell times, this calculation can be performed while the list is being processed;
the entered dwell times are observed. With very short dwell times, calculation of the
hardware settings increases the dwell time for the initial processing cycle; the entered
value is only observed from the second processing cycle onwards. In this case a mes-
sage appears to inform the user that there is a deviation between the current and set
dwell times. No further calculations are required after the first run through a list. The
current dwell times will definitely no longer deviate from the set dwell times.
The list is processed from the beginning to the end of the list (modes "Auto", ("Exter-
nal") "Single", ("External") "Step").
List Mode Dialog
To open the "List Mode" menu, select "RF/A Mod > Configure > List Mode" or use the
MENU key under "RF/A Mod".
The menu is used to activate/deactivate the operating mode List, to create, select and
activate the lists, and to select the trigger mode and the dwell time.
General Settings
State - List Mode
Activates/deactivates the List mode. The currently selected list is processed.
In case of a new or modified list, the necessary hardware settings are automatically
determined on activation of the list mode. The data determined in this way is stored
along with the list and is available whenever the list is used again.
This means that when activating the list mode, the system checks whether any hard-
ware settings are present. If so, the list is started immediately, but if not they are auto-
matically determined (the list is learnt).
A "Learn List Mode Data" button is available for deliberately activating list learning.
Note:
Activating the list mode automatically deactivates all sweeps. During list mode
the frequency and level indications do not display the currently set values.
RF Block