![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 279](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735279.webp)
Instrument Function
R&S
®
SMBV100A
264
Operating Manual 1176.8016.02 ─ 17
The "LF Frequency Sweep" dialog is used to configure and activate an LF frequency
sweep signal.
To open the "LF Frequency Sweep" dialog, select "RF/A Mod > Configure > LF Fre-
quency Sweep" or use the MENU key under "RF/A Mod".
The LF sweep mode is activated and the sweep mode is selected. The buttons are
used to reset the LF sweep (all sweep modes) or to execute the LF sweep ("Single"
mode).
The sweep range, sweep spacing and dwell time are set in the bottom of the section.
State
Activates the LF frequency sweep signal generation.
Note:
Activating a sweep mode automatically deactivates other sweeps and the list mode.
Remote command:
[:SOURce<hw>]:LFOutput:FREQuency:MODE
Mode
Selects the LF frequency sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
Generates a continuously repeating sweep signal immediately after
activating the sweep mode.
The sweep steps are performed atomatically, controlled by the dwell
time, see
Example:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR AUTO
SOUR:LFO:FREQ:MODE SWE
RF Block
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...