![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 643](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735643.webp)
Remote Control Commands
R&S
®
SMBV100A
628
Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:PHASe:STEP
<Step>
The command sets the step width by which the start phases of the carriers in the
defined carrier range will be incremented.
The phase settings are only valid if optimization of the crest factor is disabled
(
:SOURce:BB:ARB:MCARrier:CFACtor:MODE OFF
).
Parameters:
<Step>
float
Range:
-359.99 to 359.99
Increment: 0.01
*RST:
0
Default unit: DEG
Example:
BB:ARB:MCAR:EDIT:CARR:PHAS 90 DEG
sets a start phase of 90° for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:PHAS:STEP 1 DEG
the start phase is incremented by 1° for each carrier, i.e. the first
carrier has a start phase of 90°, the second a start phase of 91°,
etc.
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:PHASe[:STARt]
<Start>
The command sets the start phase for the individual carriers in the defined carrier
range. If the command
:BB:ARB:MCAR:EDIT:CARR:PHAS:STEP
is used to define a
step width, the phase entered here applies only to the starting carrier. The phases of
the remaining carriers are stepped up or down by the phase value specified in
the
:BB:ARB:MCAR:EDIT:CARR:PHAS:STEP
command.
The phase settings are only valid if optimization of the crest factor is disabled
(
:SOURce:BB:ARB:MCARrier:CFACtor:MODE OFF
).
Parameters:
<Start>
float
Range:
0 to 359.99
Increment: 0.01
*RST:
0
Default unit: DEG
Example:
BB:ARB:MCAR:EDIT:CARR:PHAS 90 DEG
sets a start phase of 90° for the carriers in the carrier range.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
<Step>
The command sets the step width by which the starting power of the carriers in the
defined carrier range will be incremented.
SOURce Subsystem
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...