![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 609](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735609.webp)
Remote Control Commands
R&S
®
SMBV100A
594
Operating Manual 1176.8016.02 ─ 17
Example:
AWGN:CNR 10
sets a carrier/noise ratio of 10 dB.
Manual operation:
See
" Carrier/Noise Ratio / Signal/Noise Ratio "
[:SOURce<hw>]:AWGN:DISP:MODE
<Mode>
Selects the display mode to Bseband or RF.
Parameters:
<Mode>
RF | BB
*RST:
RF
Example:
AWGN:DISP:MODE RF
sets the display mode to RF
Manual operation:
See
[:SOURce<hw>]:AWGN:ENRatio
<EnRatio>
In Additive Noise mode, sets the ratio of bit energy to noise power density.
Depending on the selected reference mode either the noise power (
AWGN:POW:RMOD
NOISe
) or the carrier power (
AWGN:POW:RMOD CARRier
) is adjusted to the set E
b
/N
0
ratio.
●
AWGN:POW:RMOD CARRier
the noise power is adjusted to the set E
b
/N
0
ratio and the carrier power is kept con-
stant.
●
AWGN:POW:RMOD NOISe
the carrier power is adjusted to the set E
b
/N
0
ratio and the noise power is kept con-
stant.
For
Digital Standard
signals, the bit rate used for calculation of E
b
/N
0
value from C/N
value is defined with command
SOUR:AWGN:BRAT.
For
Custom Digital Mod
signals the bit rate used for calculation of E
b
/N
0
value from
C/N value is determined by the selected standard (
SOURce:BB:DM:STANdard
) and
cannot be changed.
Parameters:
<EnRatio>
float
Range:
-50 to depends on hardware
Increment: 0.01
*RST:
15.84
Default unit: dB
Example:
AWGN:ENR 10
sets a ratio of bit energy to noise power density of 10 dB
Manual operation:
See
SOURce Subsystem