Instrument Function
R&S
®
SMBV100A
121
Operating Manual 1176.8016.02 ─ 17
In this dialog you can perform internal calibration routines.
The remote commands required to define these settings are described in
Adjust All
Performs all available internal calibration routines of the instrument.
Remote command:
Adjust Synthesis
Performs all adjustments which affect the frequency.
Remote command:
:CALibration<hw>:FREQuency[:MEASure]?
Adjust Level
Performs all adjustments which affect the level. The acquired correction values
improve the settling time and the signal quality.
Remote command:
:CALibration<hw>:LEVel[:MEASure]?
Adjust I/Q Modulator Current Frequency
Starts the adjustment for the I/Q modulator for the currently set frequency, I/Q swap,
and baseband gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q
imbalance and quadrature.
Adjustment for only the set frequency is considerably faster than adjustment across the
entire frequency range. This is possible with "Full Range".
Remote command:
:CALibration<hw>:IQModulator:LOCal?
Adjust I/Q Modulator Full Range
Starts the adjustment for the I/Q modulator for the entire frequency range. The I/Q
modulator is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Remote command:
:CALibration<hw>:IQModulator:FULL?
General Instrument Settings